Annals of the American Thoracic Society

Rationale: Central venous pressure (CVP) can be estimated by ultrasound of the inferior vena cava (IVC), but imaging the IVC is sometimes challenging. The femoral vein is easily imaged by ultrasound and might therefore provide an alternate target for estimating CVP.

Objectives: To assess femoral vein diameter (FVD) measured by ultrasound imaging for estimating CVP.

Methods: We prospectively measured CVP and FVD in 97 patients. Receiver operating characteristic curves were used to assess the ability of FVD to predict specific CVP values: less than 10 mm Hg, less than 8 mm Hg (low CVP), and greater than 12 mm Hg (high CVP). Interobserver variability of FVD measurement was assessed in 20 patients.

Measurements and Main Results: There was moderate correlation between FVD and CVP (r = 0.66, P < 0.001). FVD less than or equal to 0.8 cm was the best predictor of CVP < 10 mm Hg, with an area under the curve (AUC) of 0.894 and a 95% confidence interval (CI) of 0.82 to 0.97. FVD less than or equal to 0.7 cm performed best for predicting low CVP (AUC = 0.97; 95% CI, 0.94–0.99) and FVD greater than or equal to 1.0 cm for high CVP (AUC = 0.80; 95% CI, 0.72–0.89). However, FVD greater than or equal to 1.2 cm had the greatest specificity (94%) for high CVP. Interobserver variability in FVD measurements was 8.3 ± 7.2%.

Conclusions: The results of this exploratory study suggest that the accuracy of FVD measured by ultrasound imaging for estimating CVP is comparable to that which has been reported for ultrasound measurement of IVC diameter. FVD may provide an alternative approach when the IVC is difficult to image. Additional studies on other cohorts of patients are warranted to validate our proposed FVD cutoff values for predicting low and high CVP.

Ultrasound of the inferior vena cava (IVC) has been used to provide a noninvasive estimation of central venous pressure (CVP) (16). The IVC is sometimes difficult to image due to poor acoustic windows or obstructing medical devices, and it takes longer to learn how to correctly image the IVC than more superficial vessels. An alternative noninvasive method for estimating CVP that is simple, rapid, readily available, and easily mastered would be desirable.

The femoral vein is relatively superficial and anatomically predictable. Previous studies have shown a high degree of correlation between pressure within the common iliac vein and superior vena cava (79). The purpose of this study was to assess use of femoral vein diameter (FVD) for estimating CVP.

This was a prospective, single-center, cross-sectional study conducted between November 2012 and February 2013 in the medical intensive care unit of Hennepin County Medical Center, Minneapolis, Minnesota. Written informed consent was obtained from the patient or surrogate decision maker. The hospital’s human subjects in research committee approved the study protocol.

To be eligible for study, patients had to be 18 years of age or older and have a central venous catheter in place for transducing CVP. All patients were mechanically ventilated at the time of study and without clinical evidence of active expiration or increased intraabdominal pressure. In sequential order, the right FVD and CVP were measured with patients supine and completely flat. Ultrasounds were obtained with a SonoSite M-Turbo (SonoSite Inc., Bothell, WA) and a vascular transducer (5–10 MHz, 38-mm linear array).

The right common femoral vein was initially identified at the inguinal ligament crease using established technique (10). Scanning caudally, the great saphenous vein takeoff was identified at the anterior-medial aspect of the common femoral vein (Figure 1B). The FVD was measured as the maximal anterior-posterior dimension of the common femoral vein using the leading edge technique, just caudal to the point where the great saphenous vein was no longer visible (Figures 1C and 1D). Compression was used to confirm that the vessel was venous and to exclude the presence of a deep vein thrombus (11). Because there was no detectable cardiac or respiratory variation in FVD diameter, measurement could be made without regard to the respiratory cycle.

After zeroing the system, the CVP was obtained from a two-channel (ECG lead II and CVP) paper strip recording with the transducer at the midaxillary position. The recorded CVP value was measured at end-expiration and end-diastole.

In a subset of 20 patients, two investigators (R.J.C. and D.R.W.) measured FVD to assess interobserver variability. Each investigator was blinded to the result of the other, and the two measurements were obtained a few minutes apart.

Descriptive statistics were used to summarize patient characteristics and study measurements. Linear regression was used to assess the correlation between FVD and CVP. To assess the accuracy of FVD for predicting specific CVP values (see below), we constructed receiver operating characteristics (ROC) curves and calculated test characteristics including sensitivity, specificity, and positive and negative predictive values. Area under the curve (AUC) and corresponding 95% confidence interval (CI) were calculated and compared using Wilcoxon statistic as described by Hanley and McNeil (12). Parameters for standard error of area assumed a nonparametric distribution.

We examined the ability of FVD to predict three different CVP values: less than 10 mm Hg, less than 8 mm Hg, and greater than 12 mm Hg. We selected CVP less than 10 mm Hg because this cut-point value was used in an earlier study that examined several noninvasive methods for predicting CVP (6). The cut-point values of less than 8 mm Hg and greater than 12 mm Hg were used to define “low CVP” and “high CVP,” respectively (13).

Results are expressed as mean ± SD. A P value of less than 0.05 was considered to indicate statistical significance. All reported P values are two-sided. All calculations were computed with the aid of SPSS v21.0 and Microsoft Excel 2013.

We enrolled 97 patients (mean age, 59 ± 15 years; 54 men). Table 1 illustrates demographic and clinical features of the study population. With the exception of a single patient who had a peripherally inserted central catheter, CVP was measured from catheters inserted in the internal jugular vein. The FVD was obtained in all patients. The mean CVP and FVD of the 97 study patients were 10.8 ± 3.3 mm Hg and 0.94 ± 0.25 cm, respectively.

Table 1. Demographic and clinical features

VariablesN = 97
Male sex54 (55)
Age, yr59 ± 15
Body mass index29 ± 9
ICU characteristics 
 MAP, mm Hg71 ± 13
 Tidal volume, ml/IBW7.8 ± 1
 FiO245 ± 12
 PEEP, cm H2O5.6 ± 2
Primary diagnosis 
 Septic shock56 (60)
 Congestive heart failure12 (13)
 Cardiac arrest8 (8)
 Toxic overdose7 (7)
 Gastrointestinal bleeding4 (4)
 Seizure3 (3)
 Stroke2 (2)
 COPD exacerbation1 (1)
 Diabetic ketoacidosis2 (2)

Definition of abbreviations: COPD = chronic obstructive pulmonary disease; IBW = ideal body weight; ICU = intensive care unit; MAP = mean arterial pressure; PEEP = positive end-expiratory pressure.

Data presented as n (%) or mean ± SD.

Figure 2 shows the bivariate correlation curve of FVD to CVP. Linear regression demonstrated a moderate correlation between FVD and CVP (r = 0.66, P < 0.001).

Figure 3 shows the area under the curve (AUC) for prediction of various CVP values by FVD. For prediction of CVP less than 10 mm Hg by FVD, the AUC was 0.894 (95% CI, 0.82–0.97). Best FVD cutoff values were obtained based on the generated ROC curve. FVD less than or equal to 0.8 cm had the best testing characteristic in predicting a CVP less than 10 mm Hg (sensitivity, 77%; specificity, 95%; positive predictive value, 89%; and negative predictive value, 90%).

For predicting a low CVP (<8 mm Hg), a FVD cutoff less than or equal to 0.7 cm had the highest sensitivity (95%) and specificity (89%), with AUC of 0.97 (95% CI, 0.94–0.99). For predicting a high CVP (>12 mm Hg), an FVD cutoff of greater than or equal to 1.0 cm had the best overall testing characteristics, with sensitivity and specificity of 70% and AUC of 0.80 (95% CI, 0.72–0.89). However, a FVD cutoff of greater than or equal to 1.2 cm had the highest specificity (94%) for predicting a high CVP.

Interobserver variability in measuring the FVD was examined in 20 cases. The mean FVD for the 40 measurements (20 for each investigator) was 1.11 ± 0.23 cm. The absolute mean difference in FVD measurements was 0.08 ± 0.07 cm, with an interobserver variability of 8.3 ± 7.2%.

Several studies have reported on use of ultrasound of the IVC or internal jugular vein for estimating CVP (16, 1417). A recent study from our institution found the IVC maximal diameter to be a better predictor of CVP than either IVC collapsibility index or internal jugular vein aspect ratio in spontaneously breathing patients (6). Unfortunately, performing high-quality ultrasound of the IVC is sometimes challenging because of occlusive dressings or poor acoustic windows. In addition, respirophasic movement of the IVC can occur during ultrasound imaging, potentially introducing errors in the measurements of vessel diameter and collapsibility (18). The latter pitfall may become more important in patients with prominent respiratory efforts and when imaging is done by less experienced operators.

In the current study we examined an alternative noninvasive method for estimating CVP: measurement of FVD. Previous studies that established a high degree of correlation between right atrial and common iliac venous pressures provided a rationale for this approach (7-9). We found that there was a statistically significant relationship between FVD and CVP, with a moderate degree of correlation between the two parameters.

In both our current and previous study (6), we evaluated the performance of various ultrasound techniques at predicting a CVP less than 10 mm Hg through use of ROC curves. It appears that FVD may be comparable to IVC diameter for predicting a CVP less than 10 mm Hg in that the AUC for FVD (0.89; 95% CI, 0.82–0.97) was only slightly lower than that previously reported for IVC diameter (0.91; 95% CI, 0.84–0.98) (12). In the current study, an ROC analysis showed that a FVD less than or equal to 0.7 cm was best able to predict a low CVP (<8 mm Hg). Although an FVD greater than or equal to 1.0 cm provided the best overall testing characteristics for predicting a high CVP (>12 mm Hg), a FVD greater than or equal to 1.2 cm had the highest specificity (94%).

We found a high degree of concordance between observers in measurements of the FVD. In comparison, a recent investigation reported a much higher degree of interobserver variability for measurement of IVC diameter (19).

There are a number of limitations to our study. First, our study was primarily designed to evaluate the correlation of FVD to CVP and to generate cutoff values for FVD to provide the best testing characteristics for specific CVP values; therefore, it was not prospectively corroborated. Second, our study population was biased to a high-CVP group, because patients had been resuscitated at the time of study. This could potentially introduce error in the testing characteristics of our low-CVP group, because a lower prevalence of this subset of patients could increase the number of false positives and negatives in the study population. Third, our study population varied considerably with respect to age, sex, BMI, and height. These population-based characteristics may influence FVD (20) and could potentially influence the relationship between CVP and FVD. For example, the FVD threshold for low and high CVP might be different for petite women than large men. Fourth, because the FVD is greatest proximally and tapers toward the superficial femoral vein, there is potential for interobserver variability in measurement (21). The minimal interobserver variability of FVD measurements in our study is likely due to uniform use of the great saphenous vein take-off as the target for measurement, and measurements at other locations may be less reliable. Finally, we studied patients who were supine, on low levels of positive end-expiratory pressure, and without evidence for increased intraabdominal pressure. Change in body position, high-level positive end-expiratory pressure, and intraabdominal hypertension could potentially alter the relationship between FVD and CVP.

In conclusion, the results of this exploratory study suggest that the accuracy of FVD for estimating CVP is comparable to that reported for ultrasound of the IVC and could provide an alternative approach when the IVC is difficult to image. A low CVP (<8 mm Hg) is very likely when FVD is less than or equal to 0.7 cm, and a FVD greater than or equal to 1.2 cm is highly predictive of a high CVP (>12 mm Hg). FVD values between 0.7 and 1.2 cm appear to be less reliable for predicting either a low or high CVP. Additional studies on other cohorts of patients are needed to validate FVD cutoff values for low and high CVP.

1 . Kircher BJ, Himelman RB, Schiller NB. Noninvasive estimation of right atrial pressure from the inspiratory collapse of the inferior vena cava. Am J Cardiol 1990;66:493496.
2 . Brennan JM, Blair JE, Goonewardena S, Ronan A, Shah D, Vasaiwala S, Kirkpatrick JN, Spencer KT. Reappraisal of the use of inferior vena cava for estimating right atrial pressure. J Am Soc Echocardiogr 2007;20:857861.
3 . Stawicki SP, Braslow BM, Panebianco NL, Kirkpatrick JN, Gracias VH, Hayden GE, Dean AJ. Intensivist use of hand-carried ultrasonography to measure IVC collapsibility in estimating intravascular volume status: correlations with CVP. J Am Coll Surg 2009;209:5561.
4 . Schefold JC, Storm C, Bercker S, Pschowski R, Oppert M, Krüger A, Hasper D. Inferior vena cava diameter correlates with invasive hemodynamic measures in mechanically ventilated intensive care unit patients with sepsis. J Emerg Med 2010;38:632637.
5 . Nagdev AD, Merchant RC, Tirado-Gonzalez A, Sisson CA, Murphy MC. Emergency department bedside ultrasonographic measurement of the caval index for noninvasive determination of low central venous pressure. Ann Emerg Med 2010;55:290295.
6 . Prekker ME, Scott NL, Hart D, Sprenkle MD, Leatherman JW. Point-of-care ultrasound to estimate central venous pressure: a comparison of three techniques. Crit Care Med 2013;41:833841.
7 . Ho KM, Joynt GM, Tan P. A comparison of central venous pressure and common iliac venous pressure in critically ill mechanically ventilated patients. Crit Care Med 1998;26:461464.
8 . Nahum E, Dagan O, Sulkes J, Schoenfeld T. A comparison between continuous central venous pressure measurement from right atrium and abdominal vena cava or common iliac vein. Intensive Care Med 1996;22:571574.
9 . Joynt GM, Gomersall CD, Choi G, Chui PT. Measurement of central venous pressure via the femoral route in abdominal compartment syndrome. Crit Care Shock 2008;11:107114.
10 . Cavezzi A, Labropoulos N, Partsch H, Ricci S, Caggiati A, Myers K, Nicolaides A, Smith PC. Duplex ultrasound investigation of the veins in chronic venous disease of the lower limbs: UIP consensus document. Part II. Anatomy. Eur J Vasc Endovasc Surg 2006;31:288299.
11 . Kory PD, Pellecchia CM, Shiloh AL, Mayo PH, DiBello C, Koenig S. Accuracy of ultrasonography performed by critical care physicians for the diagnosis of DVT. Chest 2011;139:538542.
12 . Hanley JA, McNeil BJ. A method of comparing the areas under receiver operating characteristic curves derived from the same cases. Radiology 1983;148:839843.
13 . Dellinger RP, Levy MM, Rhodes A, Annane D, Gerlach H, Opal SM, Sevransky JE, Sprung CL, Douglas IS, Jaeschke R, et al.; Surviving Sepsis Campaign Guidelines Committee including the Pediatric Subgroup. Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock: 2012. Crit Care Med 2013;41:580637.
14 . Donahue SP, Wood JP, Patel BM, Quinn JV. Correlation of sonographic measurements of the internal jugular vein with central venous pressure. Am J Emerg Med 2009;27:851855.
15 . Mintz GS, Kotler MN, Parry WR, Iskandrian AS, Kane SA. Real-time inferior vena caval ultrasonography: normal and abnormal findings and its use in assessing right-heart function. Circulation 1981;64:10181025.
16 . Nakao S, Come PC, McKay RG, Ransil BJ. Effects of positional changes on inferior vena caval size and dynamics and correlations with right-sided cardiac pressure. Am J Cardiol 1987;59:125132.
17 . Simonson JS, Shiller MB. Sonospirometry: a new method for noninvasive estimation of mean right atrial pressure based on two-dimensional echographic measurements of the inferior vena cava during measured inspiration. J Am Coll Cardiol 1988;11:557564.
18 . Blehar DJ, Resop D, Chin B, Dayno M, Gaspari R. Inferior vena cava displacement during respirophasic ultrasound imaging. Crit Ultrasound J 2012;4:18.
19 . Saul T, Lewiss RE, Langsfeld A, Radeos MS, Del Rios M. Inter-rater reliability of sonographic measurements of the inferior vena cava. J Emerg Med 2012;42:600605.
20 . Fronek A, Criqui MH, Denenberg J, Langer RD. Common femoral vein dimensions and hemodynamics including Valsalva response as a function of sex, age, and ethnicity in a population study. J Vasc Surg 2001;33:10501056.
21 . Hertzberg BS, Kliewer MA, DeLong DM, Lalouche KJ, Paulson EK, Frederick MG, Carroll BA. Sonographic assessment of lower limb vein diameters: implications for the diagnosis and characterization of deep venous thrombosis. AJR Am J Roentgenol 1997;168:12531257.
Correspondence and requests for reprints should be addressed to James Leatherman, MD, Division of Pulmonary and Critical Care, G5, 701 Park Avenue, Minneapolis, MN 55415. E-mail:

Author Contributions: R.J.C. was responsible for measurements, statistical analysis, and manuscript preparation. D.R.W. was responsible for measurements in 20 patients to assess interobserver variability. J.W.L. takes responsibility for the content of the paper and contributed to manuscript preparation.

Author disclosures are available with the text of this article at www.atsjournals.org.

Related

No related items
Comments Post a Comment




New User Registration

Not Yet Registered?
Benefits of Registration Include:
 •  A Unique User Profile that will allow you to manage your current subscriptions (including online access)
 •  The ability to create favorites lists down to the article level
 •  The ability to customize email alerts to receive specific notifications about the topics you care most about and special offers
Annals of the American Thoracic Society
13
1

Click to see any corrections or updates and to confirm this is the authentic version of record